ISCUSflex is the third generation Microdialysis Analyzer for point-of-care cerebral monitoring and advanced research. This bedside analyzer uses enzymatic reagents and colorimetric measurements and is available in two versions; as part of CMA's cerebral monitoring system and as a dedicated research system. The US version is enhanced with CLIA software that will require the user authorization/password entry to run samples, including controls and changing reagents. The CLIA software also permits Pathology personnel to produce and file an authorization log as needed. Product note of US version to be published shortly. International product note is available here For more info, please contact CMA or your local distributor. |
|||||||||||||||||||||||||
Technical Information on ISCUSflex US version |
|
| Weight | 29 lbs/13 kg |
| Voltage | 100-240 V~, 50/60 Hz |
| Power consumption | 100 VA |
| Dimensions | 430 (H) x 350 (W) x 270 (D) mm |
| Type of protection | Class 1, Type B |
| Measurement principle | Kinetic enzymatic analyzer |
| Vials | Microvials, Chromacol (CMA glass) and ETH (CMA plastic) vials |
| Samples | Microdialysates |
| Sample volume used | 0.2 - 2 無/analysis (depending on analyte) |
| Min. sample volume | Sum of sample volumes per analyte + 2 無 |
| Reagent consumption | <15 無/analysis (depending on analyte) |
| Pipetting imprecision | < 2 % (0.5 無) rel. standard deviation |
| Calibration | Automatic (every 6 hrs) |
| Warm-up time | 10 minutes |
| Measuring time | 30 seconds |
| Throughput time/test | 90 seconds |
| Detector type | Single beam filter photometer |
| Light source | Class 1 LED |
| Wavelengths | 375 and 530 nm |
| Detector cell temperature | 99 慚/37 慢 |
| Detector cell | Capillary flow cell 10 mm, 2 無 |
| Assay imprecision | <4 % rel. standard deviation for Control Sample Normal |
| Assay inaccuracy | <10 % for Control Sample Normal |
Ordering Information ISCUSflex US version |
|
|
8003295 ISCUSflex Microdialysis Analyzer
|
|